GE Voluson E8 Maintenance
We recommend using a surge protector along with a dedicated power outlet when operating the Voluson E8. Probes should be disinfected with a disinfectant wipe proven not to damage the lens after every use.
Dimensions and Weight of the Voluson E8
Height: 1393 mm (54.8 in) (adjustable, maximum)
Width: 580 mm (22.8 in)
Depth: 920 mm (36 in)
Weight: 131 kg (289 lb) (no Peripherals), approx 400 lbs with packaging
Voluson E8 Specifications
Digital Beam former
1,979,578 System Processing Channel Technology
Displayed Imaging Depth: 0 – 30 cm
Minimum Depth of Field: 0 – 1 cm (Zoom, probe dependent)
Maximum Depth of Field: 0 – 36 cm (probe dependent)
Up to 274 dB Dynamic Range
GE Voluson E8 Electrical Power
Voltage: 100, 115-130, 220-240 VAC
Frequency: 50/60 Hz (+/-2%)
Power Max: 1000 VA with on-board Peripherals
Thermal Output: 3446 BTU/h
Reviews
There are no reviews yet.